1. Datalogger SK-L200 II Series
1-1. About the main unit specifications
1-2. About DATALOGGER for WINDOWS

1. Datalogger SK-L200 II Series
1-1. About the main unit specifications
Q1: Can we get the calibration with the temperature and humidity traceability certificate?
A1: Yes. Our service to customers includes providing the temperature and humidity standards. If you have a calibration request, please contact the store where you bought the instrument or our service network.
Q2: The measuring range is set to -10°C to 60°C. What is the heat resistance of the case?
A2: The case can endure heat of up to 80°C or as low as -20°C. If used outside the measuring range, the error code of Hi or Lo appears on the LCD. Be sure to use the instrument within the range of -10 to 60°C, regardless of the upper/lower temperature limit set for the case.
Q3: We are planning to use the instrument to measure the temperature of products during transport by truck. Will engine noise or continuous vibration cause a problem?
A3: If the truck travels on a normal road, the engine noise or the vibration will not be a problem. However, precaution should be taken against accidental large vibration that could damage the instrument.
Q4: How can we protect the instrument from malfunctioning due to static electricity or electrical noise?
A4: Electrical noise can be partly eliminated by placing the main body of the instrument in a metal case or wrapping it with aluminum foil. Before using the instrument in a place for the first time, be sure to check the noise level, as this will differ according to the environment.
Q5: The instrument is going to be used for continuous data logging. How long does the battery last?
A5: The battery lasts approximately 500 hours in measuring mode. This instrument can log 8100 readings at any logging interval (LCD off). For example, if the logging is set to 90 minutes, approximately 506 days (LCD off) of continuous logging is possible (assuming the instrument has two new alkaline batteries and is used at room temperature).
Q6: Does the battery life differ depending on the logging interval?
A6: Yes. Battery power is consumed faster with a shorter logging interval. The data below are our estimates assuming room temperature, normal humidity, two new alkaline batteries, and the LCD turned off. Note that the data is not guaranteed for every battery to be used, as the battery capacity differs depending on the manufacturer or environment. For a long term recording, use new batteries.
Interval Time for 8100 data Number of full data Operating time
1 sec. 2.25 hours Approx. 450 times 1,014 hours
10 sec. 22.5 hours Approx.227 times 5,113 hours
1 min. 135 hours (5.63 days) Approx.122 times 16,563 hours
10 min. 56.25 days Approx.20 times 27,749 hours
60 min. 337.5 days Approx.3 times 29,600 hours
Q7: What will happen to the data if the LCD fails during logging?
A7: The data will not be affected, but the LCD may malfunction and not properly display the operation status including the communication status, depending on the degree of LCD failure. In the worst case, data collection may be affected.
Q8: Is data collection affected if there is a machine with a high-voltage cable nearby?
A8: The instrument is hardly affected by 200-VAC cables. However, the instrument should be used as far from suspected noise sources as possible to avoid potential problems.
Q9: What will happen to the measurement data if the instrument is forcibly reset due to condensation found on the logger case during logging?
A9: This instrument should log measurement data successfully unless condensation builds up inside the indicator circuit. However, we cannot guarantee this.
Q10: I lost the cap for the communication port. What will happen to the instrument if it is used in a very dusty place without the cap?
A10: There may be a loose contact in the data signal line. The sign of trouble is failure to communicate with the computer. (The message "Connection is incorrect" appears when the connection is checked.)
Q11: Please explain the data backup of the indicator.
A11: The data remains stored even when the batteries are taken out. The data can be erased by operating the logger key or operating the "Erase" command from the computer. Be sure to download the data before erasing or starting a new measurement.
Q12: While in the logging mode, the bars and "F" are lit and the power was turned off. Why?
A12: The power of the unit is turned off to save battery power when the recorded data reaches 8100 loggings or the set end-point. When the logged data is collected from the instrument, press the POWER key to turn on the unit and then connect the unit to the computer to download the data.
Q13: Is it possible to measure a temperature exceeding 105°C?
A13: A sensor (SK-LTII-7 or SK-LTII-8) that is especially designed for high temperature is available. Please use either of them according to your propose. When using, be sure to observe the operating environment (-10 to 60°C, no condensation). Also, note that the SK-LTII-2 cannot be used for such purposes if its cable is exposed to high temperature that exceeds the temperature upper limit.
Q14: What does "Hi" on the LCD mean?
A14: It appears if the humidity falls outside the display range (10 to 99% RH).
Q15: Can I retrieve erased data from the indicator?
A15: No, there is no way to retrieve erased data. Be sure to download the data before erasing or starting a new measurement.
Q16: Can I see the measurement data while logging?
A16: Yes. Press the CALL/OUT key to view the data. Note that if the measurement data is displayed during logging, the battery is consumed faster and thus logging may be terminated halfway. If this happens, such halfway recorded data cannot be downloaded. To download, you must end recording.
Q17: What is the indicator display accuracy?
A17: It is ±(0.1 + 1 digit)°C.
Q18: How does the display accuracy change with changes in the indicator's ambient temperature?
A18: It is ±0.5°C (resolution: 0.1°C) ±1 digit (resolution: 1°C) at the operating range for the indicator (-10 to 60°C).
Total accuracy (display value) = sensor accuracy + indicator accuracy (influence of ambient temp.)
Q19: Why does the logging stop when an "Er" code is displayed?
A19: This is to protect the data recorded. When an error has occurred in the sensor, the indicator side is also affected. For this reason, recording is terminated and communication with the memory is cut off to protect the system.
Note: The "Er" code is also displayed if the sensor probe is disconnected from the indicator.
Q20: Can I use a USB connection with a USB to RS-232C serial converter cable?
A20: No. This instrument uses a unique data format designed by us. If such converter cable is used, proper communication cannot be assured.
Q21: Can I extend the USB cable length by connecting another USB cable?
A21: We guarantee successful communication only when the USB cable (1.5 meters) that comes with the instrument is used.
Q22: What type of connector is used for the communication cable?
A22: Computer side port: A plug
Indicator side port: Mini B plug*
*: Standard USB connector
Q23: Is it possible to save the data recorded in the indicator onto a commercially available external USB memory?
A23: No. The data cannot be directly downloaded to external USB memory. This instrument transfers data by the command received from the computer. The logged data should be downloaded to the computer and saved there.
Q24: Is it possible for a computer to perform multiple real-time communications?
A24: No. One computer can communicate with only one Datalogger.
Q25: The sensor probe accuracy is mentioned on the Test Result Report; what is the accuracy when it is connected to the indicator?
A25: The indicator and the sensor probe have compatible accuracy. For the accuracy of this Datalogger, it should be considered as "the total accuracy". The total accuracy consists of the "display accuracy of the indicator" inclusive of the compatible accuracy and the "accuracy of the sensor probe".
Q26: The instrument was being used outdoors, and it turned off without my turning it off. Why?
A26: It was probably shut down because the indicator shorted due to condensation. Since the indicator is not waterproof, it is highly recommended not to use it in a place subject to condensation or waterdrops.
What will happen if condensation forms on the sensor during logging (error sate is obtained):
(1) L200 series Datalogger : Logging continues.
(2) L200TII : Logging ends and "Er" appears.
(3) L200THIIα : Logging continues.
Q27: Is it possible to use this instrument by placing the sensor in an incubator?
A27: No. It must not be used in an environment subject to cultured germs or bacteria. If it is used in such a place, the humidity sensing membrane will be damaged, resulting in measurement errors.
Q28: Is it possible to use the indicator in an atmosphere with ambient conditions that are not room temperature and humidity? For example: low temperature of 5°C, high temperature of 40°C, high humidity of 75% RH.
A28: Yes. You can use it, but be careful not to allow condensation to form. If it does, the unit may malfunction. Note that regular maintenance should be performed if the instrument is used in such a severe environment.
Q29: Is it possible to use the instrument in a dusty environment?
A29: This instrument is not dust-proof, so we cannot guarantee the product if it is used in such an environment.
Q30: Is it possible to use the instrument by installing the indicator on a wall with magnet? What about the influence of the magnet?
A30: Avoid using the instrument with a strong magnet that can secure the indicator to the wall. Doing so may affect the operation of the unit.
Q31: Is it possible to use the instrument in a pressurized or decompressed environment?
A31: It is possible, but the accuracy cannot be assured. The instrument has been tested by using a humidity generator of approx. 1 bar (1013 hPa). The accuracy of the instrument is assured when it is used under the pressure of 1 bar (1013 hPa).
TOP ▲

1-2. About DATALOGGER for WINDOWS
Q1: I was not able to install the program.
A1:
(1) Was there any software program running at the same time with the installation?

If another program is running, installation may fail.
Turn off any other programs.

(2)

Did you take the correct installation steps?

If the installation ends in a state described below, the installation may not be completed.
Perform the installation procedure from the beginning again.
  1. The order of installation was wrong.
  2. The USB driver program was not installed to the USB port. (This installation must be done the first time the indicator is connected to the computer.)

Note: If you reinstall the program, be sure to uninstall the old program before reinstallation.

A. Confirming the installation (for Windows2000 and Windows XP)

 ①Checking if the USB driver is properly installed
  1. Connect the datalogger to the computer.
    From Control Panel, select System, Hardware and Device Manager.
  2. Click [+] of USB (Universal Serial Bus) controllers to indicate the list.
  3. Confirm that [SK-L200 SERIES Composite Device] is in the list.
    If there is the "! " mark is displayed next to [SK-L200 SERIES Composite Device], you have to update the driver.
    If [SK-L200 SERIES Composite Device] is not listed, it may be displayed in the Device Manager as [Unknown device].
    You have to reinstall the driver.
  - To update the driver,
  1. Point [SK-L200 SERIES Composite Device] or [Unknown Device] with your mouse and right click to display and click [Update Driver].
  2. Select [No, not this time] and click [Next] to continue.
  3. Select [Install the software automatically] and click [Next] to continue.
  4. If the Warning message is displayed, click [Continue anyway].
  5. The message [Installation was completed] is displayed. Click [Finish].
  6. Confirm that the update was completed.
    If there is no "! " mark is displayed next to [SK-L200 SERIES Composite Device] in "USB (Universal Serial Bus) controller ", it means that the installation was successful.
 ②Checking if the USB driver has been installed to the USB port
  1. Connect the datalogger to the computer.
  2. From Control Panel, select System, Hardware, Device Manager and Port (COM & LPT).
    If you see "! " mark in [SK-L200 SERIES Serial Port Drivers (WDM) (COMx)] or Driver is not displayed, it may be displayed in Device Manager as [Unknown device].
    You have to reinstall the driver.

B. Confirming the installation (for Windows Vista/7/8/8.1)

 ①Checking if the USB driver is properly installed
  1. Connect the datalogger to the computer.
    • Windows Vista:
      From Control Panel, select System & Maintenance and Device Manager.
    • Windows 7:
      From Control Panel, select System & Security and Device Manager.
    • Windows 8/8.1:
      In desktop, move the cursor to the right and select Settings, Control Panel,Hardware & Sound, Device & Printer and Device Manager
  2. Click [→] of USB (Universal Serial Bus) controllers to indicate the list.
  3. Confirm that [SK-L200 SERIES Composite Device] is in the list.
    If there is the "! " mark is displayed next to [SK-L200 SERIES Composite Device], you have to update the driver.
    If [SK-L200 SERIES Composite Device] is not listed, it may be displayed in the Device Manager as [Unknown device].
    You have to reinstall the driver.
  - To update the driver,
  1. Point [SK-L200 SERIES Composite Device] or [Unknown Device] with your mouse and right click to display and click [Update Driver Software].
  2. Click [Browse my computer for driver software].
  3. Click [Browse] and select [USB] that has been deployed when the software "DATALOGGER for Windows " was installed.
  4. Put a check mark on [Include subfolders] and click [Next].
  5. The message "Installation was completed " is displayed.
  6. Confirm that the update was completed.
    If there is no "! " mark is displayed next to [SK-L200 SERIES Composite Device] in "USB (Universal Serial Bus) controller", it means that the installation was successful.
 ②Checking if the USB driver has been installed to the USB port
  1. Connect the datalogger to the computer.
  2. From Control Panel, select System, Hardware, Device Manager and Port (COM & LPT).
    If you see "! " mark in [SK-L200 SERIES Serial Port Drivers (WDM) (COMx)] or Driver is not displayed, it may be displayed in Device Manager as [Unknown device].
    You have to reinstall the driver.

(3)

Was the installation cancelled halfway by the "Warning message"?

If that is the case, the installation is incomplete. Install the program again.
Please note that we verify it is safe to ignore the "Warning message".
Therefore, click [Continue Anyway] to continue the installation.


(4)

Was there a message saying "Path or file not found"?

This message appears if Power Users or User tries to install a program in Windows 2000.
Be sure to log in as an Administrator to install a program.

Note: The Windows user account "Administrator" is responsible for the security against the external threat. Ask your network administrator before log in as an administrator. To change your account status, refer to the Help file or User Manual of your computer.


(5)

Is the OS correct?
  • The software v.5.11E:
    for Windows XP (32 bit) and Windows2000 (SP4 or better)
  • The software v.5.2E:
    for Windows Vista (32 bit)
  • The software v.6.2E:
    for Windows 7/8/8.1 (32 bit and 64 bit)
Q2: How can I update my "DATALOGGER for Windows"?
A2: The latest version of "DATALOGGER for Windows" is available from our website. Download your copy for free.
Select the version appropriate for your computer.
Before updating, do not forget to uninstall the older version of "DATALOGGER for Windows".
Download from this.
Q3: Is the software program available for Windows 3.1, MS-DOS or Windows 95/98/NT/ME?
A3: No.
Q4: The message "Connection is incorrect." is displayed.
A4:
(1) Is the power to the Datalogger turned off?

When checking for the connection to the computer, turn on the unit (indication on).


(2)

Is the version of Datalogger for Windows correct for your computer?
  • The software v.5.11E:
    for Windows XP (32 bit) and Windows2000 (SP4 or better)
  • The software v.5.2E:
    for Windows Vista (32 bit)
  • The software v.6.2E:
    for Windows 7/8/8.1 (32 bit and 64 bit)

Update it to the latest version correct for your computer if it is not.
For how to update the software version, refer to Q2.


(3)

Are there two versions of software installed on the same computer?

Uninstall the older version.


(4)

Are you using a USB to RS-232C serial converter cable?

This instrument does not work with a converter cable.


(5)

Is the USB cable properly connected?

Check if the USB connector is properly inserted.


(6)

Have you set the serial port (COM port number) in the software?

The serial port number set in "Select the Serial Port" must be the same as the port number shown on the computer connected. Check the port number connected and set the same number in the software.

Checking the port number connected

  1. Select Port (COM & LPT) while connecting the datalogger with your computer.
    • Windows 2000 (SP4 or better)/XP:
      From Control Panel, select System, Hardware, Device Manager and Port (COM && LPT).
    • Windows Vista:
      From Control Panel, select System & Maintenance, Device Manager and Port (COM & LPT).
    • Windows 7:
      From Control Panel, select System & Security, Device Manager and Port (COM & LPT).
    • Windows 8/8.1:
      From Control Panel, select Hardware & Sound, Device & Printer, Device Manager and Port (COM & LPT).
  2. Confirm the port number.
  3. In [SK-L200 SERIES Serial Port Drivers (WDM) (COMx)], the COMx becomes port number.
    Set it in "Setting the Serial Port of the software.


* If "!" mark is seen next to [ SK-L200 SERIES Serial Port Drivers (WDM) (COMx)], update the driver.
For how to update the driver, refer to A1-(2) above.


(7)

Has the installation of software been successful?

If not, finish the installation. For how to update the software, refer to A1-(2) above.


(8)

It may be restricted by your computer system.

For details, refer to the instruction manual of the computer or ask the computer manufacturer.
  1. Ports are set to be "Disable" by the "Power Saving" function.
  2. Ports are not available because they are being used by other devices (modem, for example).
If more than one computer is available, connect the product to another computer and check if they can communicate with each other.
If more than one SK-L200 series datalogger is available, connect another datalogger to the computer and check if they can communicate with each other.
Q5: Part of the screen is hidden from view. Not all the screen is displayed.
A5: Select lower DPI of your computer display.
Select Start, Control Panel, Display, Setting, Advanced and General.
If the problem still persists, change the Windows font size to smaller.
Select Start, Control Panel, Display, Appearance and Font size.
Q6: How many Dataloggers can I connect to read all recorded data at the same time?
A6: The recorded data from up to 10 Dataloggers can be read (v. 5.11E or later). Also, four recorded data can be displayed with graphs in overlapped.
Q7: Is it possible to erase unnecessary data?
A7: Yes. You can save or print only the specified range of data and erase the rest of the recorded data.
Q8: Can I read the data created by the older version of the software with the newer version?
A8: Yes. The data saved by the older version can be opened from the latest version. However, the data saved by the latest version cannot be opened from the older version.
Q9: What is a CSV format?
A9: The CSV format is a file format readable by commercial spreadsheet software. (Excel, for example) The data recorded without setting a date and time can only be saved in CSV format.
Q10: Why is my graph divided in some cases?
A10: It happens if the processing speed is too slow for the graph drawing. When saving data, broken graphs may be displayed. This is not a problem, and the saved data is safe. Data saving is done in a specified interval.
Q11: Is there a software program for viewing data on the server?
A11: No, no software available.
Q12: The software failed in starting up.
A12: Click Start, Programs and SK-SATO to turn on the software. If it fails, the shortcut link may be broken.
A user other than an Administrator may have installed the program.
Create a shortcut to the program file (.exe) and place it on the desktop.
For a stable operation, we recommend that you uninstall and reinstall the program. To change your account status to an Administrator, refer to the Help file or User Manual for your computer.
Q13: There is no "SK-SATO APPLICATON" in Programs.
A13: The software has not been properly installed. Install it again.
Q14: Can the system automatically recognize the USB port when the Datalogger is connected to it?
A14: No. When you change the USB port to be connected, be sure to edit the "Select the Serial Port" in the software program.
Q15: When installing, a message saying "The folder is invalid. Check if there is a folder the program can be installed" appears.
A15: Your computer does not meet the system requirements for the software program. Use a computer that meets the requirements.
Q16: A message saying "No new hardware was detected" appears.
A16: Copying and installing of the USB driver are not complete. Install the necessary software including the USB driver.
For how to check if the software has been installed, refer to A1-(2) above.
Q17: The graph cannot be pasted on an Excel file with the command "Paste Graph to Clipboard".
A17: The software DATALOGGER for Windows supports the Windows clipboard, but DATALOGGER for Windows does not support the Office programs (Word, Excel, etc.) clipboard. It is possible to paste the screen on the Excel fie with the command "Paste Screen to Clipboard", but not a graph.
To copy and paste a graph on an Office file, use the following key combinations. Copy to the clipboard (Ctrl + c) and then paste (Ctrl + v)
Q18: The start date and time were preset, but the logging did not start.
A18: The logging could have been cancelled because the battery power became lower than specification (LoBAT) while in standby for preset start.
Always use two new batteries for logging.
Make sure the batteries have enough power left. Note that the lithium batteries may seem to recover power if they are not used for some time.
Q19: A message saying "No date and time have been set. Data is saved in CSV format" appears.
A19: This message appears if data logging was done without setting the date and time.
The date and time setting is erased once the battery is removed for replacement. Check that the date and time are properly set when logging starts.
Note: The Datalogger for Windows is not equipped with the data alteration function so that the software can protect the quality and reliability of the logged data. The DATALOGGER for Windows determines the logged data without the date and time is unreliable and allows it to be saved only in CSV format.
The CSV format is a file format readable by commercial spreadsheet software (Excel, for example).
Q20: The logging option cannot be changed.
A20: The logging option cannot be changed while there are logged data (bars displayed on the lower LCD). Erase the logged data before changing a logging option.
Note that there is no way to retrieve the erased data, and so be sure to download data before erasing.
Q21: I cannot down load the logged data.
A21:
(1) Is the communication between DATALOGGER for Windows and the indicator active?

Does the message "Correction is incorrect" appear when the "Connect" command is used? If the message appears, installation of the software may have been unsuccessful or the serial port has not been selected in the software.
Refer to "A1" and "A 4" on this list.


(2)

Was there a message saying "Downloading failed (checksum error)"? This error occurs if there is an error (missing data, etc.) in the logged data. Check the following points.
  1. Are you using the latest version of DATALOGGER for Windows?
    If not, update it to the latest version that is compatible to your PC.
  2. Was any other software program running while downloading?
    If other program is running, turn it off and start downloading again.
  3. Is "Power Management" active?
    If it is so, downloading can be cancelled, returning in missing data.
(3) Did the message "Realtime transmission is in progress" appear? Check the following points.
  1. Is the OUT mark is lit on the indicator?
    If yes, turn off the display by pressing the CALL/OUT key.
  2. Are you using the latest version of DATALOGGER for Windows?
    If not, update it to the latest version.
(4) Was there a message "Data verification error"?
This error occurs if there is an error (missing data, etc.) in the logged data. Check the following points.
  1. Are you using the latest version of DATALOGGER for Windows?
    If not, update it to the latest version.
  2. Was any other software program running while downloading?
    If other program is running, turn it off and start downloading again.
  3. Is the computer power is controlled by the "Power schemes"?
    If it is so, downloading can be cancelled, returning in missing data.
Q22: Is it possible to edit the display range of X axis and Y axis in the graph?
A22: The graph axis setting can only be done by the software. Unfortunately, users are not allowed to modify them.
Q23: Is it possible to combine the Y axes of three data lines?
A23: No. The graphs are plotted automatically by the software.
If any advanced graph settings should be required, we recommend you should save as a CSV file and customize the graphs with a commercial spreadsheet program (Excel, for example).
Q24: Is it possible to combine two data lines and display it?
A24: Please use the "Overlap" command of the software to display two data lines at the same time.
Q25: When "Logger Information" or "Readout Only" is selected, the display is too dark to view.
A25: Change the screen setting of the software.
Select the Screen Properties, Design and Colors in this order. Check if the color is gray. The dark screen is caused by the gray color. Select other color.
Q26: Is it possible to output a waning on the by using the "Warning" command of the software?
A26: There are two types of alarms available.
(1) Visual alarm output
The temperature (humidity) at the lower screen is lit in red.
(2) Audio alarm output
By specifying the audio file, warning can be heard.
Q27: What is "Abnormal pulses detected", which is one of the conditions that causes the Datalogger to stop logging?
A27: Logging ends if the temperature or humidity data are judged as "Er". The cause includes the following.
(1) The sensor is disconnected from the indicator.
(2) The sensor wiring is broken or short circuited.
(3) There is a fault (breakage) in the cable.
Note: For (2) and (3) above, the behavior of instrument is subject to the model type.
Q28: I set the logging interval to "15 minutes", but it showed as "1 sec" in the downloaded data.
A28: This error occurs with the version 5.0 of DATALOGGER for Windows. Use the latest version.
For how to download the latest version of the software, refer to Q2 on this list.
Q29: Is there any date or time data in each measurement data?
A29: No. Date and time data are not included in each measurement data. The data are in our original format.
Q30: Is it possible to download and measure in real time by using my original program?
A30: Yes. For interface, use the USB driver bundled with the product.
Note that we cannot guarantee any proper operation or our support if you use your original program.
Q32: The numerical values are overlapped in the printed data.
A32: Check the time setting of the computer. If it is in 12h format, for example 10:00 AM/10:00 PM, the overlapping error occurs.
Q33: The date is incorrectly displayed in CSV format.
A33: Check the date setting of the computer.
If it is in Japanese area, the date display error may occur.
Q34: The "preset start date and time" and the "date and time logging started" in the downloaded data are different.
A34: Are you using the "Repeatedly" logging option? For the "Repeatedly" option, the "date and time logging ended" must be typed by the user. Be sure to type it correctly; otherwise, there will be difference between the "preset start date and time" and the "date and time logging started".
Q35: Is the DATALOGGER for Windows in compliance with the part 11 of the FDA, USA?
A35: No.
The software cannot be altered on the screen. The saved data is in binary format, and thus it is somehow alterable. We do not guarantee the software is completely non-alterable.
Q36: The year's worth of logged data must be analyzed. Is it possible to combine all data?
A36: No. However, it is possible to combine the data on the commercial spreadsheet software (Excel, for example) if the data are saved in CSV format.
Q37: Is it possible to record the measured data while communicating in real time?
A37: By using the software, it is possible to measure in real time while saving data onto the computer. Note that if the number of stored data lines exceeds 8,100, measurement will stop automatically